TY700Series

TY710 TY720

CE

● Highest Precision

  • Basic accuracy 0.020%
  • True RMS measurement

    ● New Functions for Electronic and Industrial Use
  • Simultaneous measurement and display of DC
    and AC
  • Switchable RMS/MEAN detection※
  • Low-pass Filter (LPF)※
  • AC50mV range※
  • Lo-Power resistance measurement※
    ※TY720 only

    ● Full Support for Data Management
  • A large-capacity logging-mode memory
     TY720: 10000 data
     TY710: 1000 data
  • USB communication application software
     (Optional DMM communication package)

    ● Safe Design and Wide Temperature Range forHarsh Use
  • Conforms to safety standards EN61010-1
     1000V CAT.Ⅲ 600V CAT.Ⅳ
  • Safety shutters for preventing erroneous insertion of
     test leads into current measurement terminals
  • Operating temperature range -20 to +55℃
  • Storage temperature range -40 to +70℃

    ● Closed Case Calibration

    ● 3-year Warranty

  • General Specifications

    Measurement Functions:DC voltage AC voltage DCV+ACV DC current AC current DCA+ACA resistance frequency temperature capacitance
    duty cycle decibel calculation continuity check diode test low-power resistance (TY720 only)
    For AC voltage/current RMS/MEAN detection can be switched (TY720 only).
    For AC voltage/current the low-pass filter can be turned on/off (TY720 only).
    Additional Functions:Data hold/auto hold/peak hold (TY720 only) range hold maximum/minimum/average values
    resistance capacitance zero relative and percentage value calculation manual-mode memory logging-mode memory
    auto power off backlight (white LED)
    Display:5-digit LCD: 7-segment
    Digital display: Main display; [50000] counts
    Sub-display ; [50000] counts
    Bar graph display: 51-segment
    Polarity indicator: “-” appears automatically when the polarity is negative
    Overrange indicator: “OL”
    Measuring Rate:6 times/sec (Frequency: 1 time/sec Capacitance: max. 0.03 times/sec (50mF) Resistance: 4 times/sec)
    Bar graph display: 15 times/sec
    Operating Temp. and Humidity:-20 to 55℃; 80% RH or less (no condensation) 40 to 55C: 70% RH or less
    Storage Temp. and Humidity:-40 to 70℃; 70% RH or less (no condensation)
    Temperature Coefficient:Add the accuracy 0.05/℃ to the basic accuracy at a temperature within -20 to 18℃ and 28 to 55℃.
    Power Supply:Four AA (R6) dry cells
    Battery Life:Approx. 120 hours (for continuous DC voltage measurement with alkaline cells)
    Withstanding Voltage:6.88kV for 5 seconds (between input terminals and casing)
    Dimensions:Approx. 90(W) x 192(H) x 49(D) mm
    Weight:Approx. 560g (including batteries)
    Compliance with Standards:Safety EN61010-1 EN61010-031 1000V CATⅢ 600V CATⅣ pollution level 2 indoor 2000m max. above sea level
    EMC: EN61326-1 Class B EN55011 Class B Group 1

    Basic Accuracy & Ranges

    TY710TY720
    DCV:0.02% + 50.02% + 550.000 mV to 1000.0 V
    ACV:RMS:0.7% + 300.4% + 3050.000 mV to 1000.0 V
    ACV:MEAN:
    n/a
    1.0% + 3050.000 mV to 1000.0 V
    DCV+ACV:1.0% + 100.5% + 105.000 V to 1000 V
    DCA:0.2% + 20.2% + 2500.00 μA to 10.000 A
    ACA:RMS:1.0% + 100.75% + 20500.00 μA to 10.000 A
    ACA:MEAN:
    n/a
    1.5% + 20500.00 μA to 10.000 A
    DCA+ACA:1.5% + 101.0% + 10500.00 μA to 10.000 A
    Continuity Check:Buzzer Sounds at 100 ± 50 Ω or less Range:500.0 Ω
    Resistance:0.1% + 20.05% + 2500.00 Ω to 50.000 MΩ
    LOW-power Resistance:
    n/a
    0.2% + 35.000 kΩ to 5.000 M
    Diode Test:1.0% + 2
    1.0% + 22.4000 V
    Temp.※:1.0% + 1.5 deg.C1.0% + 1.5 deg.C-200 to 1372℃
    Capacitance:1.0% + 51.0% + 55.000 nF to 50.00 mF
    Frequency:0.02% + 10.02% + 12.000 Hz to 99.99 kHz
    Duty Cycle:±1%±1%10 to 90%
    ※Temperature probe:Type K thermocouple sensor ( optional )