AC/DC Leakage Current Measurements
Model 730 (CE)
FEATURES
  • High sensitive for low range leakage current
  • Suitable for measurement of 4-20mA DC controlled circuit
  • DC mV analog signal output
  • Lowest influence from Magnetization and Terrestrial magnetism
  • Wide measuring ranges - up to DC 1000mA and AC10A
  • Sleeping current measurements for automotive electronics circuit
  • Jaw opening capability 30mmφ
Catalog PDF
SPECIFICATIONS
Safety standard:IEC61010-1 IEC61010-2-032
CATⅡ 600V or CATⅢ 300V
Measuring function:AC/DC leakage current
Measuring method:Clamp type CT
Jaw opening capability:30mmφ
Measuring ranges:DC 100mA/1000mA AC 100mA/1000mA/10A (45Hz-65Hz)
A/D conversion:Dual slope integration method
AC conversion:Average sensing rms reading
Display:Max. 2000 count on LCD with annunciators
Over range indication:"OL" mark on LCD
Data hold indication:"DH" mark on LCD
Zero adjustment:For DC current range by "0 ADJ" switch
Sampling rate:1 time/sec. for DC and 6 times/sec. for AC
Low battery indication:"Battery" mark on LCD
Analog signal output:DC 100mV full scale to each range
(Output impedance: less than 10kΩ)
Limitation of circuit voltage:Less than AC/DC 500V
Operating temperature:0~50℃ < 85%RH (without condensation)
Storage temperature:-10~60℃ < 70%RH (without condensation)
Withstanding voltage:AC 3700V/1 minute between CT and outer case
Auto power off:Approx. 10 minutes after last switch operation
(Auto power off function can be released by the switch)
Power supply:LR6 AM-3 or AA size battery×4
Current consumption:Approx. 9mA (Approx. 200 hours continuous)
Dimension:
Main unit:78(W)×155(H)×32(D)mm Approx. 280g
CT (CTP-30DC):33(W)×170(H)×24(D)mm
1.2m lead Approx. 165g
Accessories:Carrying case Batteries Instruction manual
Option:Cable for recorder
Accuracy:(23℃±5℃ less than 85%RH)

DC Current

RangeMeasuring RangeResolutionAccuracy
100mA0.1~±99.99mA0.01mA±1%rdg±10dgtt
1000mA1~±300mA0.1mA±1%rdg±10dgt
±300.1~±700mA±2%rdg±10dgt
±700.1~±999.9mA±3%rdg±10dgt
  • Influence of terrestrial magnetism : Less than ±2.0mA
  • Influence of magnetization : Less than ±2.0mA by DC 1.5A on/off
  • Influence of CT opening and closing: Less than ±1.0mA
  • Max. input current : DC 1.5A

AC Current

RangeMeasuring RangeResolutionAccuracy
100mA0~99.99mA0.01mA±2%rdg±10dgt(50/60Hz)
1000mA0~999.9mA0.1mA±2%rdg±10dgt(50/60Hz)
10A0~9.999A0.001A±1%rdg±10dgt(50/60Hz)
  • Max. input current : AC 20A